EMG-guided percutaneous placement of cement-augmented pedicle screws for osteoporotic thoracolumbar burst fractures
2017
Background: Percutaneous techniques have increasingly gained popularity in recent years. The application of technological innovation, including neuromonitoring techniques, has the potential to increase the safety and efficacy of these procedures. Methods: Thirty patients suffering from osteoporotic dorsolumbar burst fracture were prospectively enrolled in this study. The patients underwent percutaneous fenestrated pedicle screw fixation augmented with polymethylmethacrylate (PMMA) injection. A novel surgeon-dedicated neuromonitoring device was used in order to increase the safety and the accuracy of the screw insertion. A second group of 30 patients who did not undergo neuromonitoring durin…
A new tool to improve pedicle screw placement accuracy in navigated spine surgery: A monocentric study
2021
Objective Navigated instrumented spine surgery is burden by a low but significant screw mispositioning risks, respectively, for the 2D imaging system from 15 to 40% and, for the 3D imaging system, ranging from 4.1 to 11.5%. The primary objective of this study was to demonstrate the efficacy of a new "screw-like" tool in order to further decrease pedicle screws mispositioning rate during vertebral navigated spine surgery. Materials and methods Between January and June 2019 an initial case series of 18 patients were enrolled. All patients underwent a pedicle screw fixation, both in thoracic (Th10-Th12) and lumbosacral (L1-S1) spine, using O-arm (Medtronic Navigation, Louisville, Colorado) and…